Indicator: 0.2 ~5.0%
Measuring Range:
+ 0.05 ~ 0.1%, Number of Pump Strokes 2
+ 0.1 ~ 2.0%, Number of Pump Strokes 1
+ 2.0 ~4.0%, Number of Pump Strokes 1/2
Sampling Time: 1 minute per pump stroke
Colour Change: White → Pale brown
Reaction Principle: 5CO + I2O5 + H2S2O7 → I2
Coefficient of Variation: 10% ( 0.1 ~ 0.5%), 5% (0.5 ~ 2 %)
Nr. of tests: 10
Pump: GV-100S/110S
Details
Catalog
Manual
Performance
Measuring Range |
0.05 to 0.1% |
0.1 to 2.0% |
2.0 to 4.0% |
|
Number of Pump Strokes |
|
Correction Factor |
|
Sampling Time |
1 minute per pump stroke |
Detecting Limit |
0.01% (n=2) |
Colour Change |
White → Pale brown |
Reaction Principle |
5CO + I2O5 + H2S2O7 → I2 |
Coefficient of Variation |
10% (for 0.1 to 0.5 %), 5% (for 0.5 to 2 %) |
Shelf Life |
3 Years |
Corrections for temperature & humidity |
Unnecessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences (NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Acetylene |
0.1% |
+ |
Blackish brown |
Olefins |
2/3 time |
+ |
Pale brown |
Butane |
0.3% |
+ |
Pale brown (whole layer) |
Propane |
10% |
+ |
Pale brown (whole layer) |
Hexane |
0.1% |
+ |
Pale brown |
Other substance measurable with this detector tube
Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
Calibration gas generation |
High pressure gas cylinder method |
TLV-TWA |
TLV-STEL |
Explosive range |
25 ppm(2014) |
- |
12.5 to 74% |