Indicator: 1 to 14 ppm
Measuring Range:
+ 0.2 to 0.5 ppm, Number of Pump Strokes 2
+ 0.5 to 14 ppm, Number of Pump Strokes 1
+ 14 to 35 ppm, Number of Pump Strokes 1/2
Sampling Time: 30 seconds per pump stroke
Colour Change: Pink → Yellow
Reaction Principle: C5H5N + H2SO4 → C5H5N - H2SO4
Coefficient of Variation: 10% (for 0.5 to 4 ppm), 5% (for 4 to 14 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Details
Catalog
Manual
Performance
Measuring Range |
0.2 to 0.5 ppm |
(0.5) to 14 ppm |
14 to 35 ppm |
|
Number of Pump Strokes |
|
Correction Factor |
|
Sampling Time |
30 seconds per pump stroke |
Detecting Limit |
0.1 ppm (n=2) |
Colour Change |
Pink → Yellow |
Reaction Principle |
C5H5N + H2SO4 → C5H5N - H2SO4 |
Coefficient of Variation |
10% (for 0.5 to 4 ppm), 5% (for 4 to 14 ppm) |
Shelf Life |
3 Years |
Corrections for temperature & humidity |
Temperature correction is necessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Aromatic amines |
|
+ |
Yellow |
Aliphatic amines |
|
+ |
Yellow |
Ammomia |
|
+ |
Yellow |
Hydrazine |
|
+ |
Yellow |
Other substance measurable with this detector tube
Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
4-Methyl pyridine |
0.75 |
1 |
0.38 to 10.5 ppm |
Calibration gas generation |
Diffusion tube method |
TLV-TWA |
TLV-STEL |
Explosive range |
1 ppm(2014) |
- |
1.8 to 12.4 % |