Indicator: 0.25 to 5 ppm
Measuring Range:
+ 0.25 to 5 ppm, Number of Pump Strokes 2
+ 5 to 11 ppm, Number of Pump Strokes 1
Sampling Time: 2 minutes per pump stroke
Colour Change: White → Yellow
Reaction Principle:
+ CCl4 + I2O5 + H2S2O7 → COCl2,
+ COCl2 + (CH3)2NC6 H4CHO→ (CH3)2NC6 H4CHCl2 + CO2,
+ (CH3)2NC6H4CHCl2 + (C6H5)2NH(White) → Reaction product (Yellow)
Coefficient of Variation: 10% (for 0.25 to 1 ppm), 5% (for 1 to 5 ppm)
Nr. of tests: 5
Pump: GV-100S/110S
Details
Catalog
Manual
Performance
Measuring Range |
0.25 to 5 ppm |
5 to 11 ppm |
|
Number of Pump Strokes |
|
Correction Factor |
|
Sampling Time |
2 minutes per pump stroke |
Detecting Limit |
0.04 ppm (n=2) |
Colour Change |
White → Yellow |
Reaction Principle |
CCl4 + I2O5 + H2S2O7 → COCl2 COCl2 + (CH3)2NC6 H4CHO→ (CH3)2NC6 H4CHCl2 + CO2 (CH3)2NC6H4CHCl2 + (C6H5)2NH(White) → Reaction product (Yellow)
|
Coefficient of Variation |
10% (for 0.25 to 1 ppm), 5% (for 1 to 5 ppm) |
Shelf Life |
1 Year |
Corrections for temperature & humidity |
Unnecessary |
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Hydrogen chloride |
|
No |
No |
Chlorine, Bromine |
3/10 |
+ |
No( 0.4ppm) |
Methyl bromide |
2.5times |
+ |
Pale yellow ( 2.5ppm) |
1,1,1-Trichloroethane |
14 ppm |
+ |
No( 17ppm) |
Other substance measurable with this detector tube
Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
Chloropicrin |
1.1 |
2 |
0.28 to 5.5 ppm |
Calibration gas generation |
Diffusion tube method |
TLV-TWA |
TLV-STEL |
Explosive range |
5 ppm(2015) |
10 ppm(2015) |
- |