For full functionality of this site it is necessary to enable JavaScript.

EiUK SWG100 BIOEX 固定气体分析仪

General

Sampling points: 1-4

Display: 3.5¨ TFT color graphical display, back light, window protected

Keyboard: Tactile keypad password protected menu

Industrial Interfaces/alarms/communications/Data logging

Analogue: Up to 40 x 4-20 mA output/input analogue channels (self powered), individually configurable for each gases, with hold last reading function and delay resume function (user selectable)

Alarms: Up to 20 user configurable alarm which trig a relay output (free contact 24VDC/5A) on above/below condition

System Alarm: 1 system alarm

Digital: Modbus RTU as standard, optional ethernet, profinet, profibus dp and web enabled remote access

Data logging: Internal extensive data logging function with user define logging interval- Data export to SD card

Event logging: System monitoring with extensive event logging of faulty conditions to SD card for trouble shooting

Sample

Gas inlet: Stainless steel gas fittings with 1/8” ID thread

Cooler: Integrated Peltier cooler with continuos condensate draining pump

Filter: Teflon particulate filter, Internal Viton hosing

Flow: Monitored and regulated flow 40...60 l/h

Pressure: Sample Inlet Pressure -100 mbar trough 300 mbar, auto regulated

Safety

Rating: II 3G Ex nA IIA T1 GC certified for use in Zone II areas

Norms: Comply to EN650079-15 and RL 94/9/EG

Case Sniffing: Internal cabinet gas sniffer with alarm and power supply cut off

Flow: Stainless steel flow orifice flow restrictor

Physical

Dimension/Weight: 700 x 600 x 210 mm (H x W x D) wall mounting, 45 KG

IP RATE: IP 65 Stainless steel cabinet

Notes: * Accuracy of the calibration gas not included. When calibrated with-in its recommended service interval