For full functionality of this site it is necessary to enable JavaScript.

EiUK PLB 157 微槽温度校准器 (-50 ºC ~ 160º C)

Temperature Range: 

@ 20ºC: -38C to 160 C

@ 0ºC: -50 to 160°C

Stability: 

As Liquid Bath: ± 0.01 °C

As Dry block: ± 0.01 °C

Uniformity as Liquid Bath: 

Axial: ± 0.02 °C @ -10 °C

: ± 0.02 °C@ 0°C

: ± 0.01 °C @ 120 °C

Radial: ± 0.01 °C @ -10 °C

: ± 0.01 °C@ 0°C

: ± 0.02 °C @ 120 °C

Accuracy: 

with Internal Sensor: ± 0.15 °C

with External Sensor SRS (*): ± 0.08 °C   ( TC version only)

Display : 

Set Point: -50 to 160°C

Resolution: user-selectable 0.1 , 0.01 C

Units: user-selectable C, F,K

Calibration Volume: 

As Liquid Bath: 44.5 x 185 mm

As dry block: 45 x 170 mm ( dry inserts optional)

Heating time: ambient to 120°C: 25 min ( include time to stability)

Cooling time: ambient to -10°C : 25 min (include time to stability)

Thermocouple Input (TC version): 

Type: K,J,N,R,S,E,T,B

Range: 0 to 1300 °C

Accuracy: ± 1°C Include CJC error

Resolution: user-selectable 0.1 , 0.01 °C

RTD’S input ( TC version only): 

Type: Pt 100 (3/4 wires)

Range: -100 C to 700 °C

Accuracy: ± 0.3 °C

Resolution: user-selectable 0.1 , 0.01 °C

General: 

Computer interface: RS 232 as standard ( UBS adapter supplied as option) and EIUK software included

Switch Test: Yes

Ramp: Yes fully programmable

Weight: 10.2 Kg

Dimensions: 160x3350x365 mm

Power Requirement: 115 o 230 V – 50/60 Hz

Power: 300 VA

Notes: * Require SRS sensor and special UKAS matching calibration. Long term drift not included

Details


  • 质量承诺
  • 正品保修
  • 送货到家
  • 交易简单化

注册收新闻 - 获得优惠活动的机会