Specifications
Mass Range: 1-300 amu
Ion Source: High conductance closed ion source
Filaments: Independently replaceable twin Tungsten filaments
Mass Filter: Double filter (1-inch "RF only" pre-filter with 4-inch main filter)
Detector: Dual (Faraday and secondary electron multiplier)
Maximum Analyzer Operating Pressure: 1e^-3 Torr at the ion source inlet (standard), higher pressure optional
Minimum Detectable Partial Pressure: <2e^-11 Torr (for total pressures <1e^-4 Torr on baseline inlet)
Minimum Detectable Concentration: <15ppb specified with Argon or Nitrogen
Mass Stability: Better than ±0.1 amu over 8 hours
Resolution: Better than 10% valley between peaks of equal height throughout the mass range
Base Pressure Range: Ion source at maximum pressure at 1 mTorr, 10 mTorr or 100 mTorr
Process Pressure Range: Ion source at maximum pressure at 10 mTorr, 100 mTorr, 500 mTorr, 1.5 Torr, 5 Torr, 10 Torr, 40 Torr, 100 Torr or 700 Torr
Mounting Flange: DN35CF (70mm/2.75 inches OD) Conflat® flange
Vacuum Hardware: 60 l/s turbomolecular pump with high conductance analyzer housing, inlet system, UniBloc™ inlet, automated vacuum control (RVC) completely interlocked and integrated
Backing Vacuum System:
Standard: Connected to the tool foreline pump via Surge Protect assembly with a KF16 fitting
Optional: Independent chemically resistant, dry diaphragm pump with KF16 fitting for connecting the exhaust to a suitable scrubber system
Bakeout Temperature and Bakeout Jacket: Included for 180°C bakeout
Operating Temperature: Electronics: 10-40°C
Weight: Control Module: 1.7 kg
Mechanical Support: Optional stands and brackets are available
Mobile RGA Platform: Optional RGA trolley to improve versatility (footprint 18x24 inches, 455x604mm)
Pneumatics: 60-80 psig CDA
Power Requirements: 88-264 VAC, 47/63 Hz, 600 Watts
Maximum Operating Conditions: 80% RH (non condensing)
LED Status Indication: Interlock status, filament emission, SEM, power and communications
I/O Capability: 4 analog inputs and 2 outputs (plus 1 dedicated gauge input). Optional support for a large number of both analog and digital inputs and outputs, including relay control
Other Facilities: Leak check headset socket, external filament trip socket, instrument reset
Software: Process Eye® Professional fully network compatible control platform generating under 32bit or 64bit Microsoft® Windows® xP, Vista, Server 2008 or Windows 7* (*recommended)
Communication: Ethernet CAT-5e
PC Requirements: Intel® Pentium IV® or AMD Athlon xP 1.2GHz, 1GB RAM, 120 GB hard drive, dependent upon total number of sensors on the computer and the operating system in use. Multi-sensor installation may require higher specifications.
Simultaneous Multi-Sensor: Process Eye Professional client/server configuration offers flexible multi-sensor operation.
Shipping Weight: 44 lbs (20 Kg)
Compliance: Compliant to RoHS Directive 2002/95/EC (control unit)
For specific range please check the attachment below!