Indicator: 5~ 100 ppm
Measuring Range:
+ 2.5 ~ 5 ppm, 2 Number of Pump Strokes
+ 5 ~ 100 ppm, 1 Number of Pump Strokes
Sampling Time: 1 minute per pump stroke
Colour Change: Yellow → Red
Reaction Principle: 3CH3CHO + (NH2OH)3H3PO4 → H3PO4, H3PO4 + Base → Phosphate
Coefficient of Variation: 10% (5 ~ 20 ppm), 5% (20 ~ 100 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Store the tubes: 10ºC or below
Details
Catalog
Manual tube
Manual
Performance
Measuring Range |
2.5 to 5 ppm |
5 to 100 ppm |
|
Number of Pump Strokes |
|
Correction Factor |
|
Sampling Time |
1 minute per pump stroke |
Detecting Limit |
0.25 ppm (n=2) |
Colour Change |
Yellow → Red |
Reaction Principle |
3CH3CHO + (NH2OH)3H3PO4 → H3PO4 H3PO4 + Base → Phosphate |
Coefficient of Variation |
10% (for 5 to 20 ppm), 5% (for 20 to 100 ppm) |
Shelf Life |
2 Years |
Corrections for temperature & humidity |
Unnecessary |
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. |
Possible coexisting substances and their interferences (NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Ammonia |
|
 |
No |
Acrolein |
3/5 times |
+ |
Red |
Acetone |
2 times |
+ |
Red |
Methyl ethyl ketone |
6 times |
+ |
Red |
MIBK |
10 times |
+ |
Red |
Calibration gas generation |
Diffusion tube method |
TLV-TWA |
TLV-STEL |
Explosive range |
- |
C 25 ppm |
4 to 60% |