Indicator: 0.2~5ppm
Measuring Range:
+ 0.1 ~ 5 ppm, Number of Pump Strokes 5
+ 5 ~ 40 ppm, Number of Pump Strokes 1
Sampling Time: 1.5 minute per pump stroke
Colour Change: Yellow → Reddish Brown
Reaction Principle: 3HCHO + (NH2OH)3H3PO4 → H3PO4, H3PO4 + Base → Phosphate
Coefficient of Variation: 10% (for 0.1 to 0.5 ppm), 5% (for 0.5 to 5 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Details
Catalog
Manual
Manual tube
Performance
Measuring Range |
|
Number of Pump Strokes |
|
Correction Factor |
|
Sampling Time |
1.5 minutes per pump stroke |
Detecting Limit |
0.05 ppm (n=5) |
Colour Change |
Yellow → Reddish Brown |
Reaction Principle |
3HCHO + (NH2OH)3H3PO4 → H3PO4 H3PO4 + Base → Phosphate |
Coefficient of Variation |
10% (for 0.1 to 0.5 ppm), 5% (for 0.5 to 5 ppm) |
Shelf Life |
3 Years |
Corrections for temperature & humidity |
Temperature correction is necessary |
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Aldehydes |
|
+ |
Reddish brown |
Ketones |
|
+ |
Reddish brown |
Acid Gases |
|
+ |
Red |
Organic acids |
|
No |
No |
Other substance measurable with this detector tube
Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
Diisobutyl ketone |
5.8 |
4 |
0.58 to 29 ppm |
Methaldehyde |
0.65 |
3 |
0.065 to 3.25 ppm |
Propionaldehyde |
7.6 |
1 |
0.76 to 38 ppm |
Cyclohexanone |
by scale |
1/2 |
10 to 470 ppm |
Benzaldehyde |
by scale |
1 |
4 to 92ppm |
Calibration gas generation |
Diffusion tube method |
TLV-TWA |
TLV-STEL |
Explosive range |
- |
C 0.3 ppm (2014) |
7.0 to 73% |