Indicator: 50 to 500 ppm
Measuring Range:
+ 2.5 to 25 ppm, Number of Pump Strokes 2 to 10
+ 25 to 500 ppm , Number of Pump Strokes 1
+ 500 to 1000 ppm, Number of Pump Strokes 1/2
Sampling Time: 1 minutes per pump stroke
Colour Change: White → Pale yellow
Reaction Principle: 2PH3 + 6HgCl2 + 3H2O → Hg3P2 3HgCl23H2O+6HCl
Coefficient of Variation: 10% (for 2.5 to 100 ppm), 5% (for 100 to 500 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Detail
Catalog
Manual
Performance
Measuring Range |
2.5 to 25 ppm |
25 to 500 ppm |
500 to 1000 ppm |
|
Number of Pump Strokes |
|
Correction Factor |
|
Sampling Time |
1 minute per pump stroke |
Detecting Limit |
1 ppm (n = 10) |
Colour Change |
White → Pale yellow |
Reaction Principle |
2PH3 + 6HgCl2 + 3H2O → Hg3P2 3HgCl2 3H2O+6HCl |
Coefficient of Variation |
10% (for 2.5 to 100 ppm), 5% (for 100 to 500 ppm) |
Shelf Life |
3 Years |
Corrections for temperature & humidity |
Unnecessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Arsine |
1/3 time |
+ |
Brown |
Hydrogen selenide |
1/5 time |
+ |
Brown |
Hydrogen sulphide |
1/5 time |
+ |
Dark brown |
Calibration gas generation |
High pressure gas cylinder method |
TLV-TWA |
TLV-STEL |
Explosive range |
0.3 ppm (2014) |
1 ppm (2014) |
- |