Gastec 77 快速叔丁基硫醇和二甲基硫化物测量检测管SH(CH3)3CSH(1~15 mg /立方公尺)
制造商: Gastec Model: 77 - 联系
Indicator: 1 ~ 15 mg/m^3
Measuring Range: 1 ~ 15 mg/m^3, Number of Pump Strokes 1
Sampling Time: 20 minute per pump stroke
Colour Change:
+ TBM tube: Yellow → Pink
+ DMS tube: Pink → Pale yellow
Reaction Principle:
+ (CH3)3CSH + HgCl2 → (CH3)CSHgCl + HCl,
+ HCl + Base → Chloride compound,
+ (CH3)2S + KMnO4 (Pink) → Pale yellow product
Coefficient of Variation: 10% (for 1 to 5 mg/m^3), 5% (for 5 to 15 mg/m^3)
Nr. of tests: 5
Pump: GV-100S/110S
Detector Tube |
|
||
Measuring Range |
|
||
Number of Pump Strokes |
|
||
Correction Factor |
|
||
Sampling Time | 2 minutes per pump stroke | ||
Detecting Limit | 0.2 mg/m3 (n=1) | ||
Colour Change |
|
||
Reaction Principle | (CH3)3CSH + HgCl2 → (CH3)CSHgCl + HCl HCl + Base → Chloride compound (CH3)2S + KMnO4 (Pink) → Pale yellow product |
||
Coefficient of Variation | 10% (for 1 to 5 mg/m3), 5% (for 5 to 15 mg/m3) | ||
Shelf Life | 2 Years | ||
Corrections for temperature & humidity | Temperature correction is necessary | ||
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. |
TBM Tube | |||
Substance | Concentration | Interference | Change colour by itself to |
Hydrogen sulphide | + | Pink | |
Mercaptans | + | Pink | |
DMS Tube | |||
Substance | Concentration | Interference | Change colour by itself to |
Olefins | + | Pale yellow | |
Tetrahydrothiophene | + | Pale yellow |
Calibration gas generation | TBM Tube:Diffusion tube method DMS Tube:Permeation tube method |
TLV-TWA | TLV-STEL | Explosive range |
DMS:10ppm(2014) | - | - |