Indicator: 2 to 30 ppm
Measuring Range:
+ 0.5~1ppm, Number of Pump Strokes 8
+ 1~2 ppm, Number of Pump Strokes 4
+ 2 to 30 ppm, Number of Pump Strokes 2
+ 30 to 60 ppm, Number of Pump Strokes 1
Sampling Time: 1minutes per pump stroke
Colour Change: Blue → Yellow
Reaction Principle: SO2 + BaCl2 +H2O → BaSO3 + 2HCl, HCl+Base → Chloride
Coefficient of Variation: 10% (for 2 to 10 ppm), 5% (for 10 to 30 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Detail
Performance
Measuring Range |
|
||||
Number of Pump Strokes |
|
||||
Correction Factor |
|
||||
Sampling Time | 1 minute per pump stroke | ||||
Detecting Limit | 0.1 ppm (n=8) | ||||
Colour Change | Blue → Yellow | ||||
Reaction Principle | SO2 + BaCl2 +H2O → BaSO3 + 2HCl HCl+Base → Chloride |
||||
Coefficient of Variation | 10% (for 2 to 10 ppm), 5% (for 10 to 30 ppm) | ||||
Shelf Life | 3 Years | ||||
Corrections for temperature & humidity | Unnecessary | ||||
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences (NOTE)
Substance | Concentration | Interference | Change colour by itself to |
Carbon monoxide, Nitric oxide |
No | No | |
Carbon dioxide | 100% | Unclear demarcation | Unclear demarcation |
Nitrogen dioxide | ≧ 1/1 | + | Pale purple |
Hydrogen sulphide | No | No |
Other substance measurable with this detector tube
Substance | Correction Factor | No. of pump strokes | Measuring Range |
Thionyl chloride | 0.72 | 2 | 1.44 to 21.6 ppm |