Indicator: 300-5000 ppm
Measuring Range: 300-5000 ppm, Number of Pump Strokes 1
Sampling Time: 1.5 minutes per pump stroke
Colour Change: Pale blue → Purple
Reaction Principle: CO2 + N2H4 → NH2NHCOOH
Coefficient of Variation: 10% (300 ~ 1000 ppm), 5% (1000 ~ 5000 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Details
Catalog
Manual
Manual tube
Performance
Measuring Range |
300 to 5000 ppm |
Number of Pump Strokes |
300 to 5000 ppm (n=1) |
Correction Factor |
1 |
Sampling Time |
1.5 minutes per pump stroke |
Detecting Limit |
30 ppm (n=1) |
Colour Change |
Pale blue → Purple |
Reaction Principle |
CO2 + N2H4 → NH2NHCOOH |
Coefficient of Variation |
10% (for 300 to 1000 ppm), 5% (for 1000 to 5000 ppm) |
Shelf Life |
3 Years |
Corrections for temperature & humidity |
Unnecessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences (NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Ammonia |
|
No |
No |
Carbon Monoxide |
|
No |
No |
Chlorine |
|
No |
No |
Hydrogen chloride |
|
No |
No |
Nitrogen oxides |
|
No |
No |
sulphur dioxide |
|
No |
No |
Amines |
|
No |
No |
Organic solvents |
|
No |
No |
Calibration gas generation |
High pressure gas cylinder method |
TLV-TWA |
TLV-STEL |
Explosive range |
5000 ppm(2014) |
30000 ppm(2014) |
- |