Indicator: 0.2~6%
Measuring Range: 0.02 ~12%
Sampling Hours: 0.5 to 10
Colour Change: Pale Red → Yellow
Reaction Principle: CO2 + 2KOH → K2CO3 + H2O
Coefficient of Variation: 10% (for 0.2 to 6.0 %-hr)
Nr. of tests: 10
Detail
Performance
Measuring Range | 0.02 to 12% |
Sampling Hours | 0.5 to 10 hours |
Colour Change | Pale Red → Yellow |
Reaction Principle | CO2 + 2KOH → K2CO3 + H2O |
Coefficient of Variation | 10% (for 0.2 to 6.0 %-hr) |
Shelf Life | 2 Years |
Corrections for temperature & humidity | Temperature correction is necessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences (NOTE)
Substance | Concentration | Interference | Change colour by itself to |
Ammonia | 500 ppm | No | No ( ≦ 500ppm) |
Hydrogen chloride | 300 ppm | No | No ( ≦ 300 ppm) |
Chlorine | 10 ppm | No | No ( ≦ 5 ppm) |
Hydrogen cyanide | 50 ppm | No | No ( ≦ 30 ppm) |
sulphur dioxide | 15 ppm | No | No ( ≦ 15 ppm) |
Nitrogen dioxide | 10 ppm | No | No ( ≦ 10 ppm) |
Hydrogen sulphide | 50 ppm | No | No ( ≦ 30 ppm) |
Calibration gas generation | High pressure gas cylinder method |
TLV-TWA | TLV-STEL | Explosive range |
5000 ppm (2014) | 30000 ppm (2014) | - |