Indicator: 0.5 to 10 ppm
Measuring Range:
+ 0.045 to 0.1 ppm, Number of Pump Strokes 2
+ 0.1 to 10 ppm, Number of Pump Strokes 1
+ 10 to 22 ppm, Number of Pump Strokes 1/2
Sampling Time: 2 minute per pump stroke
Colour Change: White → Yellow
Reaction Principle:
+ CCl3NO2 + I2O5 + H2S2O7 → COCl2
+ COCl2 + (CH3)2NC6H4CHO → (CH3)2NC6H4CHCl2 + CO2
+ (CH3)2NC6H4CHCl2 + (C6H5)2NH→ Reaction product
Coefficient of Variation: 10% (for 0.1 to 3.0 ppm), 5% (for 3.0 to 10.0 ppm)
Nr. of tests: 5
Pump: GV-100S/110S
Detail
Performance
Measuring Range | 0.045 to 0.1 ppm | 0.1 to 10 ppm | 10 to 22 ppm |
Number of Pump Strokes | 2 | 1 | 1/2 |
Correction Factor | 0.45 | 1 | 2.2 |
Sampling Time | 2 minutes per pump stroke | ||
Detecting Limit | 0.03 ppm (n=2) | ||
Colour Change | White → Yellow | ||
Reaction Principle |
CCl3NO2 + I2O5 + H2S2O7 → COCl2 |
||
Coefficient of Variation | 10% (for 0.1 to 3.0 ppm), 5% (for 3.0 to 10.0 ppm) | ||
Shelf Life | 1 Year | ||
Corrections for temperature & humidity | Unnecessary | ||
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences (NOTE)
Substance | Concentration | Interference | Change colour by itself to |
Ammonia | No | No discolouration | |
Carbon Dioxide | No | No discolouration | |
Carbon Monoxide | ![]() |
![]() |
No discolouration |
Ethylene Oxide | No | No discolouration | |
Formaldehyde | No | No discolouration | |
Hydrogen Cyanide | No | No discolouration | |
Bromomethane | ![]() |
+ | Yellow ( ![]() |
Phosphine | No | No discolouration | |
Sulphuryl Fluoride | No | No discolouration | |
1,2-Dichloroethane | No | No discolouration | |
Benzene | No | No discolouration | |
Toluene | No | No discolouration |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
0.1 ppm(2013) | - | - |