Indicator: 0.2~1.5ppm
Measuring Range:
+ 0.04 to 0.1 ppm, Number of Pump Strokes 10
+ 0.1 to 1.5 ppm, Number of Pump Strokes 5
+ 1.5 to 2.4 ppm, Number of Pump Strokes 3
+ 2.4 to 10 ppm, Number of Pump Strokes 1
Sampling Time: 1.5 minute per pump stroke
Colour Change: Yellow → Red
Reaction Principle: AsH3 + 3HgCl2 → As(HgCl)3 + 3HCl, HCl + Basic compound → chloride compounds
Coefficient of Variation: 10% (for 0.1 to 0.5 ppm), 5% (for 0.5 to 1.5 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Detail
Performance
Measuring Range |
|
||||
Number of Pump Strokes |
|
||||
Correction Factor |
|
||||
Sampling Time | 1.5 minutes per pump stroke | ||||
Detecting Limit | 0.02 ppm (n=10) | ||||
Colour Change | Yellow → Red | ||||
Reaction Principle | Arsine react with mercuric chloride to produce hydrogen chloride. The indicator of the tube turns to red. AsH3 + 3HgCl2 → As(HgCl)3 + 3HCl HCl + Basic compound → chloride compounds |
||||
Coefficient of Variation | 10% (for 0.1 to 0.5 ppm), 5% (for 0.5 to 1.5 ppm) | ||||
Shelf Life | 2 Years | ||||
Corrections for temperature & humidity | Unnecessary | ||||
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences (NOTE)
Substance | Concentration | Interference | Change colour by itself to |
Hydrogen chloride | No | No | |
Diborane | No | No | |
Phosphine | 1/10 time | + | Red |
Calibration gas generation | High pressure gas cylinder method |
TLV-TWA | TLV-STEL | Explosive range |
0.005 ppm (2014) | - | - |