Indicator: 0.5 to 10 ppm
Measuring Range:
+ 0.2 to 0.5 ppm , Number of Pump Strokes 4
+ 0.5 to 10 ppm, Number of Pump Strokes 2
+ 10 to 32 ppm , Number of Pump Strokes 1
Sampling Time: 2 minute per pump stroke
Colour Change: Yellow → Red
Reaction Principle:
+ CH2:C(CH3)CN+ Cr6+ + H2SO4 → HCN
+ 2HCN + HgCl2 → 2HCl+Hg(CN) 2
+ HCl+Base → Chloride
Coefficient of Variation: 10% (for 0.5 to 2 ppm), 5% (for 2 to 10 ppm)
Nr. of tests: 5
Pump: GV-100S/110S
Details
Catalog
Manual
Performance
Measuring Range |
0.2 to 0.5 ppm |
0.5 to 10 ppm |
10 to 32 ppm |
|
Number of Pump Strokes |
|
Correction Factor |
|
Sampling Time |
2 minutes per pump stroke |
Detecting Limit |
0.1 ppm (n=4) |
Colour Change |
Yellow → Red |
Reaction Principle |
CH2:C(CH3)CN+ Cr6+ + H2SO4 → HCN 2HCN + HgCl2 → 2HCl+Hg(CN) 2 HCl+Base → Chloride |
Coefficient of Variation |
10% (for 0.5 to 2 ppm), 5% (for 2 to 10 ppm) |
Shelf Life |
3 Years |
Corrections for temperature & humidity |
Unnecessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Nitriles ( C3) |
|
+ |
Red |
Acetone cyanohidrin |
|
+ |
Red |
Alcohols, Esters, Ketones |
20 ppm |
 |
No |
Aromatic hydrocarbons |
20 ppm |
 |
No |
Hydrogen chloride, Hydrogen cyanide |
|
No |
No |
Calibration gas generation |
Diffusion tube method |
TLV-TWA |
TLV-STEL |
Explosive range |
1 ppm(2014) |
- |
- |