Gastec 132TP 检测三氯乙烯气体管 Cl2C:CHCl(1 - 33 ppm)

Calibration Range: 1 - 15 ppm 
Measuring Range: 1 - 33 ppm 
Sampling (flow) Rate: 100mL/min,50mL/min
Sampling Volume: 1000mL,500mL
Sampling Time: 10 min
Colour Change: Yellow → Reddish purple
Nr. of tests: 5 tubes
Automatic Gas Sampler: GSP-300FT-2

Detail

Catalog

Specifications
Calibration Range 1 - 15 ppm
Measuring range 1 - 33 ppm
Sampling (flow) Rate 100mL/min,50mL/min
Sampling Volume 1000mL,500mL
Sampling Time 10min
Detecting Limit 0.2ppm
Colour Change Yellow → Reddish purple
Corrections for temperature/humidity Necessary for 0 to 40ºC
Unnecessary for R.H.0 to 90%
Shelf life 2 years
Accuracy CV=5-10%
(CV: Coefficient of variation = σ: standard deviation / mean value x 100)
Possibly coexisting substances and their interferences (NOTE)
Substance(s) Interference Colour change by itself
Hydrogen chloride 1/10 + Reddish purple
Chlorine 1/40 White
Vinyl chloride 1/10 + Reddish purple
1,2-Dichloroethylene 1/10 + Reddish purple
Tetrachloroethylene 1/20 + Reddish purple
1,1,1-Trichloroethane 400ppm No No
Toluene,Xylene 10ppm No
Calibration gas generation Diffusion tube method
 
TLV-TWA TLV-STEL Explosive range
10 ppm 25 ppm --

 

  • 质量承诺
  • 正品保修
  • 送货到家