Gastec 132HA 快速三氯乙烯检测管 Cl2C:CHCl(20~1300 ppm)
- 制造商: - Gastec - Model: - 132HA 
- 联系
Indicator: 50 to 500 ppm
Measuring Range: 
+ 20 to 50 ppm, Number of Pump Strokes 2
+ 50 to 500 ppm, Number of Pump Strokes 1
+ 500 to 1300 ppm, Number of Pump Strokes 1/2
Sampling Time: 45 seconds per pump stroke
Colour Change: Yellow → Reddish Purple
Reaction Principle: Cl2C:CHCl + PbO2 + H2SO4 → HCl, HCl + Basic Compound → Chloride
Coefficient of Variation: 10% (for 50 to 100 ppm), 5% (for 100 to 500 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
| Measuring Range | 
 | |||
| Number of Pump Strokes | 
 | |||
| Correction Factor | 
 | |||
| Sampling Time | 45 seconds per pump stroke | |||
| Detecting Limit | 4 ppm (n=2) | |||
| Colour Change | Yellow → Reddish Purple | |||
| Reaction Principle | Cl2C:CHCl + PbO2 + H2SO4 → HCl HCl + Basic Compound → Chloride | |||
| Coefficient of Variation | 10% (for 50 to 100 ppm), 5% (for 100 to 500 ppm) | |||
| Shelf Life | 2 Years | |||
| Corrections for temperature & humidity | Temperature correction is necessary | |||
| Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. | ||||
| Substance | Concentration | Interference | Change colour by itself to | 
| Nitric Oxide, Nitrogen dioxide | No | No | |
| Hydrogen chloride, Chlorine, Bromine | + | Reddish purple | |
| Acetone | ≦ 200 ppm | No | No | 
| 1,1,1-Trichloroethane | + | Reddish purple ( ≧ 3000 ppm) | |
| Unsaturated halogenated hydrocarbons | + | Reddish purple | |
| Aromatic hydrocarbons | ≧ 100 ppm |  | No | 
| Substance | Correction Factor | No. of pump strokes | Measuring Range | 
| 1,2-Dichloroethylene | 1.6 | 1 | 80 to 800 ppm | 
| 1,3-Dichloropropene | 0.9 | 2 | 45 to 450 ppm | 
| Calibration gas generation | Diffusion tube method | 
| TLV-TWA | TLV-STEL | Explosive range | 
| 10 ppm | 25 ppm | - | 