Indicator: 0.1 to 2.5%
Measuring Range:
+ 0.01 to 0.05%, Number of Pump Strokes 2
+ 0.05 to 2.5%, Number of Pump Strokes 1
+ 0.05 to 2.5%, Number of Pump Strokes 1/2
Sampling Time: 1 mimute per pump stroke
Colour Change: Pink → Pale blue
Reaction Principle: C2H5OH + Cr6+ + H2SO4 → Cr3+
Coefficient of Variation: 15% (for 0.05 to 0.5 %), 10% (for 0.5 to 2.5 %)
Nr. of tests: 10
Pump: GV-100S/110S
Detail
Catalog
Manual
Performance
Measuring Range |
0.01 to 0.05% |
0.05 to 2.5% |
2.5 to 7.5% |
|
Number of Pump Strokes |
|
Correction Factor |
|
Sampling Time |
1 minute per pump stroke |
Detecting Limit |
0.004% (n=2) |
Colour Change |
Pink → Pale blue |
Reaction Principle |
C2H5OH + Cr6+ + H2SO4 → Cr3+ |
Coefficient of Variation |
15% (for 0.05 to 0.5 %), 10% (for 0.5 to 2.5 %) |
Shelf Life |
3 Years |
Corrections for temperature & humidity |
Temperature correction is necessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Carbon monoxide |
|
No |
No |
Carbon dioxide |
|
No |
No |
Alcohols |
|
+ |
Pale blue |
Calibration gas generation |
Static gas dilution method |
TLV-TWA |
TLV-STEL |
Explosive range |
- |
1000ppm(2014) |
3.3 to 19% |